Design and implementation of embedded system for chl-a fluorescence detection

Detta är en Master-uppsats från KTH/Skolan för elektroteknik och datavetenskap (EECS)

Sammanfattning: Over the last decades, the effects of climate change have become increasingly evident across natural environments. Apart from other areas, climate change poses a serious threat on water quality. More specifically, it is expected that the effects of global warming around the world will severely limit our ability to control the spread and occurrence of Harmful Algal Blooms (HABs) in the future. A HAB episode is characterised by rapid proliferation of algal biomass which can cause major implications on the environment, the ecosystems, on human health, the economy, and societies overall. One way of detecting the presence of algae is to determine the chlorophyll-a concentration levels in water. This project proposes an embedded system for early algae detection in water samples by means of chl-a fluorometry. The system makes use of a constructed sensor to detect chl-a fluorescence emission. Two versions of the sensor were designed and implemented, both of which were calibrated and then tested. Afterwards, the results were presented, and the system’s performance was evaluated and discussed. Lastly, it was concluded that the performance of the system was adequate for detecting a 50 μg/L limit of chl-a concentration, however, careful testing of the site is required for field applications in samples of natural water.